MEDICAL OIL-FREE COMPRESSORS
COMMITTED TO SUSTAINABLE PRODUCTIVITY We stand by our responsibilities towards our customers, towards the environment and the people around us. We make performance stand a test of time. This is what we call Sustainable Productivity.
ROAD TO OIL-FREE Start Finish the oil-free medical world!
ROAD TO OIL-FREE Today Oil injected technology Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
CLEAN AIR IS A MUST FOR PATIENT SAFETY Two ideologies: 1. HTM/ISO take in ambient air, use any type of air compression method (oil-injected or oilfree) and ensure patient safety by means of an ultra high purification system (MED + QDT) 2. NFPA99 (US) / CSA 7396-1 (CAN) be more stringent on air quality intake and ensure patient safety by not adding contaminants during the air compression stage In normal operation both systems will provide clean air, however only oil-free solutions fully eliminate the risk of oil carry-over to the patient
ROAD TO OIL-FREE
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
TODAY OIL-INJECTED AIRPLANTS GA-MED Proven technology Using Atlas Copco s premium range of oil-injected rotary screw compressors Adapted for use in medical Airplants Incorporating Mk5 Elektronikon graphic controller Airplants complying to HTM & ISO requirements
OIL-FREE, CONTAMINANT FREE RISK FREE WHAT'S WRONG WITH OIL IN AIR? 2-3 mg/m³ = 5 LITERS / YEAR
1)UNDER-USAGE A tangible threat in medical ISO7396-1 (HTM02-01 for UK) requires 3 redundant compressors The design flow is often hugely oversized! Emulsification The addition of terminal units in areas with low usage or occupancy adds a need for additional flow capacity which is seldom needed think of medical air in ward areas for example HOW OFTEN DO YOU SEE THESE IN USE? Sometimes? Often? Occasionally? ALMOST NEVER... All this under-usage leads to creation of condensation & blockages downstream. Solutions? Oil-free (or VSD)
1)UNDER-USAGE A tangible threat in medical Other applications vulnerable to under-usage: Teaching facilities Military facilities Laboratories or R&D facilities Sterilisation facilities Other causes of under-usage: Oversizing for future expansion of a hospital adds up to 25% on top of an already oversized system IF THE AIRPLANT IS SIGNIFICANTLY UNDER-USED OIL-FREE (OR VSD) ELIMINATES THE RISK!
2)OIL CARRY OVER A tangible threat in medical ISO7396-1 (HTM02-01 for UK) requires 2 redundant Air treatment devices (MED) Performance stated at 20 C! What happens in real life?
CONSEQUENCES OF OIL A failure of any of the air treatment components could cause: Risk to patient health Require complete rebuilding of hospital pipelines Damage to expensive equipment (e.g. anesthesia machine)
3) LACK OF MAINTENANCE Causing damage to dryer components Although oil carry-over happens in medical systems, generally this does not cause harm to patients More likely the problem is contamination of components Saturated filters are often the route cause! This is often a problem with maintenance (OR LACK OF)! In particular this is a problem where there is little or no dedication to preventive or routine maintenance.
3)LACK OF MAINTENANCE Causing damage to Air treatment components Does this sound familiar in your facility? If it does then oil-free solutions offer complete piece of mind By eliminating the risk of oil damaging the purification system = TOTAL PATIENT SAFETY
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
ATLAS COPCO IS A Global market Leader in oil-free technology Active in oil-free for more than 60 years We have installed > 40,000 oil-free machines World wide We are The First manufacturer to receive ISO 8573-1 CLASS 0 certificate We offer the widest range of pressures (up to 13 bar - AQ) and FAD (2 l/s - 155 l/s), using 3 different technologies to produce 100% oil-free air: piston - tooth - water injected screw
ENVIRONMENTAL FRIENDLY FACT The removal of waste oil is expensive, difficult to handle and an environmental hazard Local regulations oblige additional investments to remove and treat contaminated condensate.
CONCLUSION THE ADVANTAGES OF OIL-FREE AIRPLANTS No risk of oil carry-over No risk of damaged dryer components No emulsification More environmental friendly Higher customer value Clean air is a choice you make choose oil free Oil-free, contaminant free risk free
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
ROAD TO OIL-FREE
OIL-FREE MEDICAL COMPRESSORS Application window We use different technologies to match customer s needs: * COMMON MEDICAL RANGE
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
LEGAL REQUIREMENTS FOR MEDICAL COMPRESSORS Medical Device Directive 93/42/EEC Due to the compressor being considered part of a Medical Device (Medical Airplant), it should: Have a risk analysis Software should be compliant with EN62304 (class A) Software should be extensively tested acc. to standard Test report should be created and available on request EMC test performed (Domestic/industrial Environment!) Be part of technical file (MDMF) for a Medical Airplant Have automatic-restart-after-power-failure activated Additional safety features based on Risk Analysis Specific test procedures during design, production, installation and commissioning Tested as a complete package Medical Airplant ISO HTM
LEGAL REQUIREMENTS FOR MEDICAL COMPRESSORS HTM 02-01, HTM2022 items in addition to ISO7396-1 General functions of compressors: individual starter units an isolator interlocked with the covers an industrial grade ampmeter a total hours counter Control functions (communicated to hospital via alarms): control circuit failed motor tripped aftercooler temperature high compressor temperature high compressor failed-to-go-on-load activation of other safety devices supplied by the manufacturers
INDUSTRIAL OR MEDICAL EQUIPMENT It is all about human beings
OIL-FREE PLANT BASED ON PISTON TECHNOLOGY LF-MED
LF-MED PRODUCT RANGE LF-MED* Single-stage oil-free reciprocating compressors designed for operating at maximum working conditions Available in 3 power ratings: 5, 7 & 10 kw Pressure of 10 bar (145 psi) and capacities from 3.1 to 15.5 l/s at 50 Hz, 7.6 to 38.6 cfm at 60 Hz Scope of supply: Power Pack (PP) Compressor block, transmission & motor * LE-MED (oil-lubricated piston technology) also available
LF-MED MAIN ADVANTAGES Based on a durable, high performance LFcompressor Always uses Y-D control for easy & fast start Supplied with advanced control system (Mk5 Elekronikon Graphic), including: Full integration with ES-MED Central Controller Easy service and diagnosis Includes medical safety features, like: Full remote monitor/control Automatic restart after voltage failure (ARAVF) Emergency Forced Local (EFL) Additional outlet temperature & FTGOL monitor Complies to the compressor requirements of medical standards: HTM 02-01 & HTM 2022 ISO 7396-1 EN 62304 (Class A software device)
OIL-FREE PLANT BASED ON TOOTH TECHNOLOGY ZT-MED & ZR-MED
ZT/ZR-MED RANGE Features: Medical software and reliability Combined inlet silencer & filter Workplace - Reduced noise level Vertical cooler set-up with integrated WSD Electronic Water Drain Environmental friendly No risk of oil contamination Variable Speed Drive (VSD) available Best energy saver
TECHNICAL SPECIFICATION: ZT-MED & ZR-MED PLANT ISO version: ZT15-ZT55 VSD (15-55 kw) Voltage variants: 400 V 50Hz Air and Water cooled (ZT/ZR) uair Max delivery by plant, l/s (2 back-up compressors; ISO7396-1) 180,0 160,0 140,0 120,0 100,0 80,0 60,0 40,0 20,0 0,0 1(+2) x ZT15 MED 1(+2) x ZT18 MED 1(+2) x ZT22 MED 2(+2) x ZT15 MED 2(+2) x ZT18 MED 2(+2) x ZT22 MED 3(+2) x ZT15 MED 3(+2) x ZT18 MED 4(+2) x ZT15 MED 3(+2) x ZT22 MED 4(+2) x ZT18 MED 4(+2) x ZT22 MED
OIL-FREE PLANT BASED ON WATER INJECTED COMPRESSORS AQ-MED ONLY ISO!
AQ-MED VSD THE BEST COMBINATION Maximizing efficiency, saving energy The premium solution AQ-MED VSD Energy savings Medical software and reliability Environmental friendly Stable pressure band Progressive start No intensity peaks Increases element reliability Advanced design No oil No lipseals No gearbox
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
SELECTION MATRIX OF OIL-FREE TECHNOLOGY Piston: LF-MED Scroll: SF-MED* Water-lubricated: AQ-MED Tooth: ZT-MED Flow of plant, l/s (l/m) 1 or 2 backups 0 53 0-3.180 5 30(44) 300 1.630(2.400) 22 282 1.320 16.920 25-154 1.500 9.240 Pros Simple to service Vessel mounted version available Suitable for remote location Compact Lowest capital cost Cons Recommend to use sound hound to reduce noise emission Efficiency Very silent Compact design All elements controlled individually giving high efficiency and savings Max ambient air +45 C Low pressure for combined med air systems VSD technology as standard Pressure up to 13 bar Efficient cooling High efficiency at large flows Most efficient cost/flow Not for HTM02-01 Large dimensions Highest capital cost Matrix gives a basic information and comparison. Technology is overlapping. In most applications multiple choices are available.
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price?
WHAT ABOUT CUSTOMER VALUE? SEE THE FULL PICTURE Risk of damage to expensive equipment Risk of additional repairs Risk of contamination Disposal (filters, oil) Oil treatment (OSC) Investment Oil-injected Oil-free CUSTOMER VALUE
OIL FREE A UNIQUE PROPOSAL Medical airplants based on best-in-class oil-free compressors No chance of oil Full compliance with requirements of medical standards Peace of mind All equipment complies with EN 62304 Lock-out spec 4 oil-free compressor technologies are possible Unique proposition Higher customer value
THE OIL-FREE WORLD Quality, Reliability, Safety Clean air is a choice you make choose oil free Oil-free, contaminant free risk free
ROAD TO OIL-FREE Today Oil injected technology Why use oil-free in medical? Which technologies can we offer? What is the difference with an industrial compressor? Finish the oil-free medical world! What are the specific features of each technology? What will be the price? How to order?
COMMITTED TO SUSTAINABLE PRODUCTIVITY.